Grabner Product |
Flashpoint |
|
MiniFlash
FLP/H Touch |
|
MINIFLASH TOUCH is a uniquely designed flash
point tester for the determination of flashpoints of
liquids and solids, using the Grabner flash point
detection method of measuring the instantaneous
pressure increase inside the continuously closed
chamber, triggered by an energy-controlled electric
arc. Heating the test chamber from the top avoids
condensation of high volatile compounds,
significantly improving the test results. |
Its continuously closed cup design and small sample volume of only 1-2 mL
offers maximum safety, avoids offensive fumes in the testing area and
reduces sample waste. The fast thermoelectric cooling at the end of a
test and the fast and easy handling of the instrument save expensive
labor time.
The full color touch-screen of the MINIFLASH TOUCH runs on Microsoft®
Windows® software and features an intuitive menu navigation that
requires no training for flashpoint testing. Along with its convenient
touch screen design, the MINIFLASH TOUCH offers an extended temperature
range from 0 to 400°C, hassle-free communication with USB, Ethernet,
LIMS and PCs and nearly unlimited storage capacity for programs and
results.
MINIFLASH determines the flash point according to ASTM D6450 and
according to the advanced ASTM D7094 standard. This procedure describes
the “Standard Test Method for Flash Point by Modified Continuously
Closed Cup (MCCCFP)”, which offers the highest repeatability and
reproducibility and an excellent correlation to the D93 Pensky-Martens
method: "there is no statistically significant bias observed between
ASTM D7094 and ASTM D93 Procedure A” as concluded during the last Round
Robin. ASTM D7094 has been approved by the ASTM committee for the Diesel
Fuel and Kerosene specifications ASTM D975, D396, D2880 and D3699.
Furthermore, test results are equivalent to other well known standards
such as Abel and TAG, and thanks to its unique design MINIFLASH can be
used for more difficult applications such as flash-point testing of
contaminated samples like diesel fuel in lube oil or gasoline in diesel
fuel and for biodiesel testing. The new MINIFLASH TOUCH analyzer also
includes improved methods for simulation of EN ISO 3679 and EN ISO 3680.
Due to true portability and rugged housing, the MINIFLASH TOUCH is also
the perfect solution for mobile labs and military applications. |
|
Features
‧Flashpoint according to ASTM D6450/SHT0768
(Flash-Point by Continuously Closed Cup) and
ASTM D7094 (Flash-Point by Modified
Continuously Closed Cup)
‧FLASHPOINT COMMITTEE STATEMENT: "There is
no statistically significant bias observed
between ASTM D7094 and ASTM D93 Procedure
A."), equivalent EN ISO 2719, excellent
correlation to ASTM D56, ISO13736, IP170
‧Approved for Diesel Fuel and Kerosene
specifications ASTM D975, D396, D2880, D3699
‧New methods for EN ISO 3679 and EN ISO 3680
simulation
‧Dilution Flashpoint Testing
‧Go / No Go Flashpoint Testing
‧Fast Screening Methods
‧Automatic ignition cleaning method
‧Maximum safety through Grabner continuously
closed cup method (Ignition Protection
Technology™)
‧1-2 ml sample size
‧No Open Flame, no offensive vapors in the
lab
‧Highest accuracy (see precision statement)
‧Extended temperature range: 0 to 400°C with
external cooling (MINIFLASH FLPH TOUCH)
‧Fully remote operable
‧Big, color touch-screen for convenient,
intuitive menu navigation
‧Touch Screen Runs on Microsoft® Windows®:
Full integration with network, LIMS and PC
‧Hassle-free communication and data transfer
via USB and Ethernet
‧USB printer support
‧Good Laboratory Practice: User Access
Control
‧Portable for Mobile and Field Use (car
battery connector)
‧Worldwide standard method for the flavors
and fragrances industry
‧Approved for the ASTM specifications of
recycled fuel and biodiesel
‧Short measuring time (depends on sample
type) due to small sample volume of 1-2 ml
‧Combustion Analysis™
‧Excellent thermoelectric heating and
cooling regulation through Peltier
Protection Technology™ |
Technical Data
‧Temperature Range
MINIFLASH FLP TOUCH 0 to 200°C (32 to
390°F)
MINIFLASH FLPH TOUCH 0 to 400°C (´32 to 750°F)
‧Power Requirements:
100/110/120/230/240 V AC 50/60 Hz, 160/210 W
‧Field Application: 12 V/8A DC (vehicle
battery)
‧W x H x D: 253 x 368 x 277 mm (10" x
14.5" x 10.9")
‧Weight: 12 kg (26 lb)
Precision Data |
MINIFLASH
ASTMD6450 |
MINIFLASH
ASTMD7094 |
Pensky-Martens
ASTM D93A |
Repeatability |
1.9°C (3.4°F) |
3.8°C (6.8°F) |
4.1°C (7.4°F) |
Reproducibility |
3.1°C (5.6°F) |
5.4°C (9.7°F) |
6.9°C (13.4°F) |
|
|
|
|