Grabner Product |
Flashpoint |
|
MiniFlash
FLP/H/L |
|
MINIFLASH
is a uniquely designed flash point series of
instruments for the automatic determination of the
flash point of liquids and solid samples according
to ASTM D6450 and the advanced ASTM D7094 standard.
This procedure describes the “Standard Test Method
for Flash Point by Modified Continuously Closed Cup
(MCCCFP)”, which offers the highest repeatability
and reproducibility and an excellent correlation to
the D93 Pensky-Martens method: "There is no
statistically significant bias observed between ASTM
D7094 and ASTM D93 Procedure A” as concluded during
the last Round Robin. ASTM D7094 has been approved
by the ASTM committee for the Diesel Fuel and
Kerosene specifications ASTM D975, D396, D2880 and
D3699.
|
Furthermore, test results are equivalent to
other well known standards such as Abel and TAG,
and thanks to its unique design, MINIFLASH can
be used for more difficult applications such as
flash-point testing of contaminated samples like
diesel fuel in lube oil or gasoline in diesel
fuel and for biodiesel testing.
Its continuously closed cup design and small
sample volume of only 1-2 mL offers maximum
safety, avoids offensive fumes in the testing
area and reduces sample waste. The fast
thermoelectric cooling at the end of a test and
the fast and easy handling of the instrument,
save expensive labor time.
Due to the true portability and rugged housing,
the MINIFLASH is also the perfect solution for
mobile labs and military applications.
The New Technology
The MINIFLASH series of instruments is using the
Grabner flash detection method of measuring the
instantaneous pressure increase inside the
continuously closed chamber due to a high energy
electric arc. The fast thermoelectric cooling
after each test prepares the instrument for the
next test in a very short time. Heating the test
chamber from the top avoids condensation of high
volatile compounds, significantly improving the
test results
|
Features
‧Complies with the latest
flashpoint standard ASTM D7094 - MCCCFP
(Flash-Point by Modified Continuously Closed
Cup) and ASTM D6450 - CCCFP (Flash-Point by
Continuously Closed Cup)
‧Excellent Correlation to ASTM D93, D56,
ISO2719 (EN/DIN 22719), ISO13736, IP170
‧Maximum Safety through Ignition Protection
Technology™: only 1-2 ml sample, no open
flame, continuously closed cup, no offensive
vapors in the lab
‧Approved by: US D.O.T. and RCRA
‧Approved by: NATO; 6630-25-145-3256,
6630-25-146-0895, 6630-01-534-1774 and the
US NAVY - NSN: 6625-01-472-6783
‧Approved for Diesel Fuel and Kerosene
specifications ASTM D975, D396, D2880, D3699
‧Approved for the ASTM specifications of
recycled fuel and biodiesel
‧The world wide standard in the flavors and
fragrances industry
‧The highest precision: details see below
(Precision data)
‧Short measuring time (depends on sample
type) due to small sample volume of 1-2 ml
‧Fast thermoelectric temperature regulation
by Peltier Protection Technology™
‧Easiest handling (instrument-user dialogue)
and fast cleaning of the sample cup
‧Wide Measuring Range: -25 to 400°C (-10 to
750°F)
‧Compact & portable for field use (car
battery connector)
‧RS232 interface and MINIWIN software (for
PC)
Technical Data
‧Recommended Temperature Range
MINIFLASH FLPL -25 to 100°C (-10 to 210°F)
MINIFLASH FLP 0 to 200°C (32 to 390°F)
MINIFLASH FLPH 10 to 400°C (50 to 750°F)
‧Power requirements: 100/110/230 V AC
50/60 Hz, 150 W
‧Field application: 12 V/8A DC (vehicle
battery)
‧W x H x D: 196 x 315 x 175 mm (7.7" x 12.4"
x 6.9")
Precision Data |
MINIFLASH
ASTMD6450 |
MINIFLASH
ASTMD7094 |
Pensky-Martens
ASTM D93A |
Repeatability |
1.9°C (3.4°F) |
3.8°C (6.8°F) |
4.1°C (7.4°F) |
Reproducibility |
3.1°C (5.6°F) |
5.4°C (9.7°F) |
6.9°C (13.4°F) |
|
|
|
|