鴻盛科儀有限公司(鸿盛科仪有限公司)
  GREEN TREE SCIENTIFIC & INSTRUMENT CO.


Grabner Product 
Flashpoint
 
 MiniFlash FLP/H/L
  MINIFLASH is a uniquely designed flash point series of instruments for the automatic determination of the flash point of liquids and solid samples according to ASTM D6450 and the advanced ASTM D7094 standard. This procedure describes the “Standard Test Method for Flash Point by Modified Continuously Closed Cup (MCCCFP)”, which offers the highest repeatability and reproducibility and an excellent correlation to the D93 Pensky-Martens method: "There is no statistically significant bias observed between ASTM D7094 and ASTM D93 Procedure A” as concluded during the last Round Robin. ASTM D7094 has been approved by the ASTM committee for the Diesel Fuel and Kerosene specifications ASTM D975, D396, D2880 and D3699.


 
Furthermore, test results are equivalent to other well known standards such as Abel and TAG, and thanks to its unique design, MINIFLASH can be used for more difficult applications such as flash-point testing of contaminated samples like diesel fuel in lube oil or gasoline in diesel fuel and for biodiesel testing.

Its continuously closed cup design and small sample volume of only 1-2 mL offers maximum safety, avoids offensive fumes in the testing area and reduces sample waste. The fast thermoelectric cooling at the end of a test and the fast and easy handling of the instrument, save expensive labor time.

Due to the true portability and rugged housing, the MINIFLASH is also the perfect solution for mobile labs and military applications.

The New Technology

The MINIFLASH series of instruments is using the Grabner flash detection method of measuring the instantaneous pressure increase inside the continuously closed chamber due to a high energy electric arc. The fast thermoelectric cooling after each test prepares the instrument for the next test in a very short time. Heating the test chamber from the top avoids condensation of high volatile compounds, significantly improving the test results

 

Features

‧Complies with the latest flashpoint standard ASTM D7094 - MCCCFP (Flash-Point by Modified Continuously Closed Cup) and ASTM D6450 - CCCFP (Flash-Point by Continuously Closed Cup)
‧Excellent Correlation to ASTM D93, D56, ISO2719 (EN/DIN 22719), ISO13736, IP170
‧Maximum Safety through Ignition Protection Technology™: only 1-2 ml sample, no open flame, continuously closed cup, no offensive vapors in the lab
‧Approved by: US D.O.T. and RCRA
‧Approved by: NATO; 6630-25-145-3256, 6630-25-146-0895, 6630-01-534-1774 and the US NAVY - NSN: 6625-01-472-6783
‧Approved for Diesel Fuel and Kerosene specifications ASTM D975, D396, D2880, D3699
‧Approved for the ASTM specifications of recycled fuel and biodiesel
‧The world wide standard in the flavors and fragrances industry
‧The highest precision: details see below (Precision data)
‧Short measuring time (depends on sample type) due to small sample volume of 1-2 ml
‧Fast thermoelectric temperature regulation by Peltier Protection Technology™
‧Easiest handling (instrument-user dialogue) and fast cleaning of the sample cup
‧Wide Measuring Range: -25 to 400°C (-10 to 750°F)
‧Compact & portable for field use (car battery connector)
‧RS232 interface and MINIWIN software (for PC)

Technical Data

‧Recommended Temperature Range
   MINIFLASH FLPL  -25 to 100°C (-10 to 210°F)
   MINIFLASH FLP    0 to 200°C (32 to 390°F)
   MINIFLASH FLPH 10 to 400°C (50 to 750°F)
‧Power requirements:  100/110/230 V AC 50/60 Hz, 150 W
‧Field application: 12 V/8A DC (vehicle battery)
‧W x H x D: 196 x 315 x 175 mm (7.7" x 12.4" x 6.9")

Precision Data

MINIFLASH
ASTMD6450

MINIFLASH
ASTMD7094

Pensky-Martens
ASTM D93A

Repeatability

1.9°C (3.4°F)

3.8°C (6.8°F)

4.1°C (7.4°F)

Reproducibility

3.1°C (5.6°F)

5.4°C (9.7°F)

6.9°C (13.4°F)